COSMOSIL BUCKYPREP-D

 
Separation of Fullerenes Metallo-fullerenes
Buckyprep-D
  • For preparative separation of derivatized fullerenes
  • For separation of derivatized fullerenes such as C60 indene used for organic thin-film solar cell
MATERIAL CHARACTERISTICS
COSMOSIL Buckyprep-D
Silicagel
high purity porous spherical silica gel
Average particle size
5µm
Average pore size
approx. 120Å
Specific surface area
approx. 300m2/g
Bonded Phase
Bonded phase
Nitro-carbazoyl group
End capping treatment
Noear-perfect
Feature
For separation of fullerene derivatives

Application of C60 Indene

Buckyprep-D offers improved separation for C60 indene, a derivatized fullerenes, that has received much attention as the n-type semiconductor material of organic thin-film solar cell.

COSMOSIL Buckyprep-D
COSMOSIL Buckyprep

Data courtesy of Yusuke Tajima, Dr. Sci.
Organic Optoelectronics Laboratory, RIKEN (Institute of Physics and Chemistry)

Column size: 4.6mmI.D.×250mm; Mobile phase: toluene; Flow rate: 1.0 ml/min.; Detection: UV 325 nm; Temperature: 30 C; Sample: C60[Indene]2 (1.0mg/ml)

Comparison with Conventional Buckyprep Column

Buckyprep-D retains derivatized fullerenes longer than C60. Therefore it is more suitable for preparative separation of derivatized fullerenes than conventional Buckyprep column.

5C18-MS-II
Buckyprep-D
Buckyprep

Column size: 4.6mmI.D.×250mm; Mobile phase: toluene; Flow rate: 1.0 ml/min.; Detection: UV 325 nm; Temperature: 30 C

Adjustment of Retention by Mobile Phase
Comparison of Each Solvent
Adding hexane or heptane to toluene increases retention on Buckyprep-D, while adding methanol or acetonitrile decreases retention. The effect of solvents on retention is the opposite of the conventional Buckyprep column.

Table. Comparison of percentage of retention ratio of each solvent on Buckyprep-D

Y-axis: Percentage of capacity factor k' when it is 100% toluene
X-axis: 10% of each solvent added into toluene

Adding Methanol Decreases Retention
Retention may decrease when a polar solvent is added on Buckyprep-D. For example, by increasing the methanol concentration to 1%, the retention dramatically decreases.

Buckyprep-D
Column size:

4.6 mm I.D. x 250 mm

Flow rate:

1.0 ml/min

Temperature:

30°C

Detection:

UV 325 nm

Sample:

[6,6]-Diphenyl-C62 Bis(Butyric Acid Methyl Ester)
(0.5mg/ml)

Inj. Vol:

5.0 µl

Solution for Too Strong Retention on Buckyprep-D
If the retention of derivatized fullerens on Buckyprep-D is too strong, shorten column length or add additives in toluene to decrease retention.

Using 50 mm Length Column

Buckyprep-D
Column size:

4.6 mm I.D. x 50 mm

Mobile phase:

Toluene

Flow rate:

1.0 ml/min

Temperature:

30°C

Detection:

UV 325 nm

Sample:

[6,6]-Diphenyl-C62 Bis (Butyric Acid Methyl Ester)
[Bis[60]PCBM] (0.50mg/ml)

Inj. Vol:

10 µl

Adding Methanol

Buckyprep-D
Column size:

4.6 mm I.D. x 250 mm

Mobile phase:

Methanol/ Toluene = 1/99

Flow rate:

1.0 ml/min

Temperature:

30°C

Detection:

UV 325 nm

Sample:

Tris [C60] PCBM

Sample cpnc:

1.0 mg/ml

Inj. Vol:

10.0 µl

Features of Each Solvent for Fullerenes Separation

Buckyprep Buckyprep Solubility of C60
(mg/ml)

Toluene

The most commonly used solvent

3.2

n-hexane

0.046

n-heptane

Weaker eluent than toluene

-

Methanol*

0.001

2-propanol*

-

Acetonitrile*

Weaker eluent than toluene. Recommended as a wash solvent for Buckyprep-D

0.018

Chlorobenzene

Stronger eluent than toluene

7.0

o-dichlorobenzene

Stronger eluent than chlorobenzene

27

1,2,4-trichlorobenzene

The strongest eluent. Recommended as a wash solvent

21.3

* These are stronger eluent on Buckyprep-D, opposite to Buckyprep.

Attention: 
Use HPLC Grade solvent or solvents after filtration or distillation. Except for alkali aqueous and strong acidic solutions, other solvents can be used (water-free pyridine and others). Depending on solvents, pay attention to high pressure caused by high solvent viscosity.

Note:
Baseline of Bukyprep-D is less stable relative to other fullerene columns. To stabilize baseline, let acetonitrile run through for 10 minutes before analysis.

Catalog # Cosmosil Buckyprep-D Column Price ($U.S.) -
09611-01 Buckyprep-D Guard Column, 4.6 mm I.D. x 10 mm $494.00
09613-81 Buckyprep-D Guard Column, 10 mm I.D. x 20 mm $1,483.00
09614-71 Buckyprep-D Guard Column, 20 mm I.D. x 50 mm $6,067.00
09646-61 Buckyprep-D Packed Column, 4.6 mm I.D. x 50 mm $896.00
09647-51 Buckyprep-D Packed Column, 4.6 mm I.D. x 250 mm $2,009.00
09648-41 Buckyprep-D Packed Column, 10 mm I.D. x 50 mm $3,554.00
09650-91 Buckyprep-D Packed Column, 10 mm I.D. x 250 mm $6,479.00
09651-81 Buckyprep-D Packed Column, 20 mm I.D. x 250 mm Call for Pricing Call

 